Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8026B: (1 year) CAL VER /5500 DATE: 10-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 38 NUMBER OF LINES: 124 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Fluke 8026B instruction manual (PN 646414 June 1982). # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 55CW8020.bmp # 55_V8020.bmp # 55_A8020.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5500A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F W 1.002 ASK+ K 1.003 HEAD EQUIPMENT SETUP 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP Allow the UUT 30 min. to stabilize. 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 80%. 1.005 DISP Turn the UUT on. 1.006 HEAD {RESISTANCE TESTS} 1.007 PIC 55CW8020 1.008 HEAD RESISTANCE TESTS: {2 kOhm Range} 1.009 5500 2 0.000kZ 0.001U CW 2.001 HEAD RESISTANCE TESTS: {200 Ohm Range} 2.002 5500 200 0.0Z 0.2U CW 3.001 5500 200 190.0Z 0.7U CW 4.001 HEAD RESISTANCE TESTS: {2 kOhm Range} 4.002 5500 2 1.900kZ 0.003U CW 5.001 HEAD RESISTANCE TESTS: {20 kOhm Range} 5.002 5500 20 19.00kZ 0.03U CW 6.001 HEAD RESISTANCE TESTS: {200 kOhm Range} 6.002 PIC 55_V8020 6.003 5500 200 190.0kZ 0.3U 2W 7.001 HEAD RESISTANCE TESTS: {2000 kOhm Range} 7.002 5500 2000 1900kZ 39U 2W 8.001 HEAD RESISTANCE TESTS: {20 MOhm Range} 8.002 5500 20 10.00MZ 0.2U 2W 9.001 HEAD {CONDUCTANCE TEST} 9.002 HEAD CONDUCTANCE TEST: {200nS Range} 9.003 5500 200 100.0nY 3U 2W 10.001 HEAD {CONTINUITY TEST AUDIBLE TONE} 10.002 RNG 2kZ 10.003 DISP Depress the AC/DC switch to activate the audible tone. 10.004 5500 0Z S 2W 10.005 EVAL Is there an {audible tone}? 11.001 DISP Disconnect the 5500A from the UUT. 11.002 DISP Apply line voltage (110V @ 60Hz) to the V Ohm S 11.002 DISP and COMMON inputs using an appropriate cable. 11.003 EVAL Is UUT {beeping at a rate of approximately 5 to 10 per sec}. 12.001 DISP Disconnect the line voltage from the UUT. 12.002 PIC 55_V8020 12.003 HEAD {DC VOLTAGE TESTS} 12.004 HEAD DC VOLTAGE TESTS: {200mV Range} 12.005 5500 200 190.0mV 0.3U 2W 13.001 5500 200 -190.0mV 0.3U 2W 14.001 HEAD DC VOLTAGE TESTS: {2V Range} 14.002 5500 2 0.000V 0.001U 2W 15.001 5500 2 1.900V 0.003U 2W 16.001 HEAD DC VOLTAGE TESTS: {20V Range} 16.002 5500 20 19.00V 0.03U 2W 17.001 HEAD DC VOLTAGE TESTS: {200V Range} 17.002 5500 200 190.0V 0.3U 2W 18.001 HEAD DC VOLTAGE TESTS: {1000V Range} 18.002 5500 1000 1000V 2U 2W 19.001 HEAD {AC VOLTAGE TESTS} 19.002 HEAD AC VOLTAGE TESTS: {200mV Range} 19.003 5500 200 190.0mV 1.2U 50H SI 2W 20.001 5500 200 10.0mV 0.3U 50H SI 2W 21.001 5500 200 10.0mV 0.5U 10kH SI 2W 22.001 HEAD AC VOLTAGE TESTS: {2V Range} 22.002 5500 2 1.900V 0.041U 10kH SI 2W 23.001 5500 2 1.900V 0.022U 5kH SI 2W 24.001 5500 2 1.900V 0.012U 2kH SI 2W 25.001 5500 2 1.900V 0.012U 50H SI 2W 26.001 HEAD AC VOLTAGE TESTS: {20V Range} 26.002 5500 20 19.00V 0.12U 50H SI 2W 27.001 5500 20 19.00V 0.12U 2kH SI 2W 28.001 5500 20 19.00V 0.22U 5kH SI 2W 29.001 5500 20 19.00V 0.41U 10kH SI 2W 30.001 HEAD AC VOLTAGE TESTS: {200V Range} 30.002 5500 200 110.0V 1.3U 2kH SI 2W 31.001 5500 200 110.0V 0.8U 50H SI 2W 32.001 HEAD AC VOLTAGE TESTS: {750V Range} 32.002 5500 750 750V 6U 50H SI 2W 33.001 5500 750 750V 6U 1kH SI 2W 34.001 HEAD {DC CURRENT TESTS} 34.002 PIC 55_A8020 34.003 HEAD DC CURRENT TESTS: {2mA Range} 34.004 5500 2 1.900mA 0.015U 2W 35.001 HEAD DC CURRENT TESTS: {20mA Range} 35.002 5500 20 -19.00mA 0.15U 2W 36.001 HEAD DC CURRENT TESTS: {200mA Range} 36.002 5500 200 190.0mA 1.5U 2W 37.001 HEAD DC CURRENT TESTS: {2000mA Range} 37.002 5500 2000 1900mA 15U 2W 38.001 END